Variable peak areas observed when injecting samples using an ACQUITY Sample Manager-FTN - WKB299612
Article number: 299612
SYMPTOMS
- Variable peak areas observed when injecting samples using an ACQUITY Sample Manager-FTN
ENVIRONMENT
- Arc HPLC System
- ACQUITY UPLC Sample Manager - Flow Through Needle
CAUSE
The syringe is not primed.
FIX or WORKAROUND
Prime the sample syringe with 30 syringe volumes.