Skip to main content
Waters

Arc FTN unable to prime sample syringe - WKB295213

Article number: 295213

SYMPTOMS

  • Air in sample syringe
  • Unable to prime sample syringe, even after switching to methanol

ENVIRONMENT

  • ACQUITY Arc Sample Manager FTN-R
  • Arc HPLC System

CAUSE

The solvent purge line was replaced with different tubing. The ID was too small and the pump could not prime.

FIX or WORKAROUND

Replace the tubing with the correct part (700010544 - ASSY, TUBE, PEEK, SOLVENT PURGE LINE).

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.