With Sciex OS, Sample Manager and Sample Organizer temperatures revert to 10 ºC when in Standby - WKB291111

Article number: 291111

SYMPTOMS

  • Temperatures are set correctly in the Instrument Method and work well during analyses
  • The Sample Manager and Sample Organizer temperatures always go to 10 ºC when the MS system switches to Standby

ENVIRONMENT

  • Sciex OS 3.0
  • Sciex OS 3.4
  • Support Layer 1.3
  • Support Layer 2.0
  • I-Class 
  • Sample Organizer
  • UPLC SM-FTN PLUS

CAUSE

Unknown

FIX or WORKAROUND

  1. Ensure that the "UAC: run all administrators in admin approval mode" setting is disabled.
  2. Go to the following file and modify the temperature as desired: C:\Program Files\Common Files\SCIEX\ClearCore\Drivers\WatersShim\Standby\StandbyMethod.xml  

 Prior to changing the file, you must reboot the PC to implement the change of the UAC setting.

This workaround should not be used in regulated environments.

ADDITIONAL INFORMATION

Also seen in Sciex OS 3.4, Waters Support Layer 2.0, DP2023 (1.79, Build 2023 03 02)