How much oil should be in the 515/1515/1525 pump? - WKB29039
Article number: 29039
ENVIRONMENT
- 515 HPLC Pump
- 1515 Isocratic HPLC Pump
- 1525 Binary HPLC Pump
ANSWER
The oil in the 515/1515/1525 pump is pushed up a tube by a gear located at the bottom of the pump. The oil then drips from the top of this tube and lubricates the rest of the mechanics. The oil level in the pump needs to be high enough to cover the gear at the bottom of the pump.