Poor system precision; RSD of peak area exceeds limit of acceptance - WKB289917
Article number: 289917
SYMPTOMS
- RSD of peak area exceeds the limit of acceptance
ENVIRONMENT
- Arc HPLC
- ACQUITY Arc
CAUSE
The purge line is bent, damaged, or broken.
FIX or WORKAROUND
Change the purge line, part 700010544 (Assy, Tube, PEEK, Solvent Purge Line).
ADDITIONAL INFORMATION
Guide the customer in how to position the tubing during usage, to prolong the shelf life of the tubing.