High pressure or system overpressure on FLR Detector - WKB277414
Article number: 277414
SYMPTOMS
- High pressure or system overpressure on FLR Detector
ENVIRONMENT
- 2475
- ACQUITY FLR
CAUSE
Blockage in the flow cell
FIX or WORKAROUND
- Flush the flow cell as described in the Operator's/Getting Started guide.
- Replace the flow cell if flushing is not successful.
ADDITIONAL INFORMATION
id277414, 2475, 2475B, A-10FL, A-30FL, UPFLRARC, UPFLRDET, UPPFLR