Skip to main content
Waters

ACQUITY UPLC Sample Manager Needle Error: "Sample Fluidics High Pressure Limit" Error - WKB27524

Article number: 27524
This article is a portion of a Troubleshooting Path.  To begin troubleshooting this issue, please select the link below.
Fix Sample Fluidics High Pressure Limit Error
A troubleshooting path for the Sample Fluidics High Pressure Limit Error
Pages: 3

SYMPTOMS

  • Instrument produces "Sample Fluidics High Pressure Limit" Error during injection cycle
  • Instrument produces "Sample Fluidics High Pressure Limit" Error during priming of wash solvents

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC I-Class
  • nanoACQUITY UPLC
  • ACQUITY Sample Manager

CAUSE

Clog inside sample needle or bent sample needle

FIX or WORKAROUND

  1. Replace sample needle.
  2. Prime wash solvents and sample syringe for at least three cycles.
  3. Calibrate X-, Y- and Zp positions.
  4. Calibrate needle z axis position.
  5. Calibrate needle seal.
  6. Calibrate needle and loop volumes.
  7. Perform sample syringe and wash syringes leak tests.
  8. Test the system with known standard.

ADDITIONAL INFORMATION

nanoACQUITY needle p/n 700002708

Standard ACQUITY PEEK needle p/n 700002644

id27524, overpressure, UPNAYXPMP, UPNSAMMGR, UPNSOLMGR

Not able to find a solution? Click here to request help.