Low sample energy on 2487/2489 detector - WKB2756
Article number: 2756
SYMPTOMS
- Sample energy is very low, but reference energy is good
- Washing the flow cell in 6N Nitric acid does not resolve issue
- Flushed unit in 100% methanol at 1 mL/min for 30 minutes but didn't resolve the problem.
ENVIRONMENT
- 2487 Dual Wavelength Absorbance Detector
- 2489 UV/Vis Detector
CAUSE
Wrong solvent like THF suspected to have been used for the testing hence causing low sample energy due to high absorbance of THF
FIX or WORKAROUND
- Replace the flow cell
ADDITIONAL INFORMATION
Similar issue can also be caused by the presence of bubble in the flow cell
id2756, 2487, 2489, 2489B, 2489SFC, A-10UV