Do I need to modify an ACQUITY UPLC system with PDA/ELS to use normal-phase solvents? - WKB27181

Article number: 27181

ENVIRONMENT

  • ACQUITY UPLC Binary Solvent Manager (BSM)
  • ACQUITY UPLC Sample Manager (SM)
  • ACQUITY PDA Detector
  • ACQUITY ELS Detector
  • Normal-phase solvents

ANSWER

Yes, modifications are required. You should install the optional THF/Hexane Kit:

  • Part 205000971 for passive CV BSM
  • Part 205000972 for i2V CV BSM

Solvents supported with this kit are:
- THF
- Hexane
- Ethyl acetate
- Acetone

Solvents NOT supported:
- Methylene chloride
- Chloroform
- Toluene
- Chlorinated solvents including trichlorobenzene
- Strong acids (>5%)

System compatibility information:

Mobile phase lines (A1, A2, B1, B2): 100% hexane/THF compatible
Plunger seal wash: 25% hexane/THF compatible

Needle wash: 100% hexane/THF compatible
Strong wash: 100% hexane/THF compatible
Weak wash: 100% hexane/THF compatible

ADDITIONAL INFORMATION

Replacing the BOS pump tubing can be very difficult on certain models; for the models/SNs listed below, it is recommended that the upgrade be performed at a Waters repair center:

  • ACQUITY UPLC (UPS model) sample managers with serial numbers between J05UPS131M and C07UPS695M
  • ACQUITY UPLC (UPA model) sample managers with serial numbers between J05UPS051M and C07UPA335M

All serial numbers after this already have the compatible needle wash tubing installed.

If a system is not fitted with the kit, up to 10% THF can be run through the system.

id27181, A-30PDA, eluent, UPAPDA, UPBINARY, UPBSM+, UPBSMIVD, UPC2PDA, UPELS, UPELSARC, UPIBSM, UPIBSM+, UPIBSMIVD, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPPBSM, UPPDA, UPPDAARC, UPPDAARCB, UPPDA-E, UPPDA-L, UPPDALTC, UPPDATC, UPPPDA, UPSMIVD, UPSMPMGR, UPSMPMGR