ACQUITY Sample Manager does not reach set point - WKB2716

Article number: 2716

SYMPTOMS

  • H-Class SMFTN is set to 5 °C, but it does not cool under 9 °C
  • No errors in ACQUITY Console error logs
  • Fan in back of SMFTN is running
  • Sample compartment not cooling to set point

 

ENVIRONMENT

  • ACQUITY H-Class Sample Manager FTN
  • ACQUITY I -Class Sample Manager FTN
  • ACQUITY I -Class Sample Manager Fixed Loop
  • ACQUITY M-Class Micro Sample Manager Fixed Loop
  • ACQUITY UPC2 Sample Manager Fixed Loop
  • ACQUITY HDX Manager
  • ACQUITY UPLC Sample Manager Fixed Loop

CAUSE

  • Frost or ice on the sample chamber heat sink

FIX or WORKAROUND

  1. Defrost Sample Manager.