ACQUITY H-Class Sample Manager FTN does not cool below 9 degrees C - WKB2716

Article number: 2716

°

SYMPTOMS

  • H-Class SMFTN is set to 5 °C, but it does not cool under 9 °C
  • No errors in ACQUITY Console error logs
  • Fan in back of SMFTN is running

 

ENVIRONMENT

  • H-Class
  • Sample Manager, FTN (SMFTN)

CAUSE

Possible condensation or frost buildup

FIX or WORKAROUND

  1. Set the temperature to 40 °C and leave it overnight.
  2. Reboot the SMFTN.
  3. Set the temperature to 5 °C.

ADDITIONAL INFORMATION

 