ACQUITY Sample Manager does not reach set point - WKB2716

Article number: 2716

SYMPTOMS

  • H-Class SMFTN is set to 5 °C, but it does not cool under 9 °C
  • No errors in ACQUITY Console error logs
  • Fan in back of SMFTN is running
  • Sample compartment not cooling to set point

 

ENVIRONMENT

  • ACQUITY H-Class Sample Manager FTN
  • ACQUITY I -Class Sample Manager FTN
  • ACQUITY I -Class Sample Manager Fixed Loop
  • ACQUITY M-Class Micro Sample Manager Fixed Loop
  • ACQUITY UPC2 Sample Manager Fixed Loop
  • ACQUITY HDX Manager
  • ACQUITY UPLC Sample Manager Fixed Loop

CAUSE

  • Frost or ice on the sample chamber heat sink

FIX or WORKAROUND

  1. Defrost Sample Manager.

ADDITIONAL INFORMATION

Instructions for defrosting ACQUITY Sample Manager.

id2716, A-30SM, LEAPPAL3, UPBSMIVD, UPIBSM, UPIBSM+, UPIBSMIVD, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPNHDXMGR, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO, UPSMIVD, UPSMPMGR, UPSMPMGR