Column Active Pre-Heater over temp fuse h/w fault on the Sample Manager-FTN - WKB271183

Article number: 271183

SYMPTOMS

  • Column Active Pre-Heater over temp fuse h/w fault
  • Resetting the sample manager does not resolve
  • Reseating the Active Preheater does not resolve

ENVIRONMENT

  • ACQUITY UPLC H-Class Sample Manager-FTN
  • ACQUITY Column Heater - Active

CAUSE

Faulty active preheater assembly

FIX or WORKAROUND

Replace the Active Pre-Heater assembly.

 

ADDITIONAL INFORMATION

What is the part number for the ACQUITY Active Preheater (APH)? - WKB7232