Pressure fluctuation with 1515 after PM - WKB270846

Article number: 270846

SYMPTOMS

  • Pressure fluctuation after PM

ENVIRONMENT

  • 1515 Isocratic HPLC Pump
  • 1525 Binary HPLC Pump
  • 515 HPLC Pump

CAUSE

Air in the pump

FIX or WORKAROUND

  1. Prime the pump to waste and ensure that a steady stream of solvent flows to waste. 
  2. Prime the pump for 15 minutes (or until the pressure reading is steady). 
  3. If pressure still fluctuates, refer to WKB117598. 