Skip to main content
Waters

Pressure fluctuation with 1515 after PM - WKB270846

Article number: 270846

SYMPTOMS

  • Pressure fluctuation after PM

ENVIRONMENT

  • 1515 Isocratic HPLC Pump
  • 1525 Binary HPLC Pump
  • 515 HPLC Pump

CAUSE

Air in the pump

FIX or WORKAROUND

  1. Prime the pump to waste and ensure that a steady stream of solvent flows to waste. 
  2. Prime the pump for 15 minutes (or until the pressure reading is steady). 
  3. If pressure still fluctuates, refer to WKB117598

ADDITIONAL INFORMATION

 

id270846, 1515, 1525, 1525U, 1525UIVD, 515, Delta, pressure ripple

Not able to find a solution? Click here to request help.