What is the minimum flow rate for the ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)? - WKB269834

Article number: 269834

ENVIRONMENT

  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)

  • ACQUITY UPLC H-Class PLUS

  • ACQUITY UPLC H-Class PLUS Bio

  • ACQUITY UPLC H-Class

  • ACQUITY UPLC H-Class Bio

ANSWER

The range of Flow rates can set depends on device's firmware.

Firmware Minimum Maximum Increments
1.5x and eariler 0.010 mL/min 2.000 mL/min 0.001 mL/min
1.60 0.010 mL/min 2.200 mL/min 0.001 mL/min
1.6x and later 0.001 mL/min 2.200 mL/min 0.001 mL/min

Maximum operating pressure of 103,421 kPa (1034 bar, 15,000 psi) at flow rates to 1 mL/min.

The maximum operating pressure decreases linearly to 53,779 kPa (538 bar, 7800 psi) at flow rates up to 2.2 mL/min.