Noisy baseline on RI detector 2414 - WKB267529
Article number: 267529
SYMPTOMS
- Noisy baseline
- Can affect area reproducibility due to baseline pickup
- Noise fails specification and SQT noise and drift test
ENVIRONMENT
- 2410 Differential Refractometer
- 2414 Refractive Index Detector
- ACQUITY UPLC RI Detector
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
- ACQUITY Arc
CAUSE
Solvent not homogeneous
Lines other than isocratic line have different solvents that slowly bleed into the flow, affecting RI signal (more of an issue on QSM and QSM-R).
Seal wash solvent also mixing with flow
FIX or WORKAROUND
Place all solvent lines into the same solvent bottle and prime thoroughly, including seal wash.
ADDITIONAL INFORMATION
Seal wash solvent affecting baseline in a cyclic way
id267529, 2414, 2414B, A-10RI, A-30SDM, UPAQSM, UPARQSM, UPPQSM, UPQSM, UPQSM, UPQSM+, UPQSM+, UPQSMBIO, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO