Unstable pressure on ACQUITY UPLC H-Class Quaternary Solvent Manager - WKB267555
Article number: 267555
SYMPTOMS
- Unstable pressure on the QSM
- Check valves, seals, and piston sonicated
- Dynamic leak test passes
- Pump ran as expected for about 10 to 15 minutes
- Pressure on primary pump dropped to 0 psi and accumulator pressure fluctuated greatly
- Sonicating i2V check valve resolves the issue
ENVIRONMENT
- ACQUITY UPLC H-Class Quaternary Solvent Manager
CAUSE
Faulty or dirty i2V check valve
FIX or WORKAROUND
- Sonicate the i2V check valve cartridge in 50:50 MeOH/water.
- If issue returns, replace the i2V check valve cartridge.
ADDITIONAL INFORMATION
id267555, checkvalve, DLT, methanol, pressure fluctuation, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO