Unstable pressure on ACQUITY UPLC H-Class Quaternary Solvent Manager - WKB267555

Article number: 267555

SYMPTOMS

  • Unstable pressure on the QSM
  • Check valves, seals, and piston sonicated
  • Dynamic leak test passes
  • Pump ran as expected for about 10 to 15 minutes
  • Pressure on primary pump dropped to 0 psi and accumulator pressure fluctuated greatly
  • Sonicating i2V check valve resolves the issue

ENVIRONMENT

  • ACQUITY UPLC H-Class Quaternary Solvent Manager

CAUSE

Faulty or dirty i2V check valve

FIX or WORKAROUND

  1. Sonicate the i2V check valve cartridge in 50:50 MeOH/water.
  2. If issue returns, replace the i2V check valve cartridge. 

ADDITIONAL INFORMATION