ACQUITY UPLC "Degasser Pressure Not Reached" error - WKB267488
Article number: 267488
SYMPTOMS
- "Degasser Pressure Not Reached" error
- Power cycle and priming do not resolve error
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY Arc
- ACQUITY UPLC H-Class
CAUSE
The degasser exhaust tube is pinched or folded, obstructing flow.
FIX or WORKAROUND
- Straighten or adjust the tubing.
- Reset the module and monitor the degasser until it reaches and maintains 0.9 psi.