2695 leaks from the check valve - WKB261399

Article number: 261399

SYMPTOMS

  • Leak from the check valve

ENVIRONMENT

  • 2695 Separations Module
  • Alliance HPLC System

CAUSE

Loose check valve housing

FIX or WORKAROUND

Tighten the check valve housing.

ADDITIONAL INFORMATION

If tightening the check valve housing does not stop the leak - replace the check valve housing - Part number # 700002332 