"Left/Right valve home h/w fault" for Column Manager - WKB251750
Article number: 251750
SYMPTOMS
- "Left/Right valve home h/w fault" shows in Console
ENVIRONMENT
- ACQUITY UPLC Column Manager
- ACQUITY Arc Column Manager
CAUSE
Failed column valve
FIX or WORKAROUND
Replace the valve:
- 700008871 - I-Class
- 700005438 - H-Class/Arc
- 700005461 - H-Class/Arc Titanium
- 700005438 - UPC2
ADDITIONAL INFORMATION