Skip to main content
Waters

"Plate missing" error appears when trying to run a sample set in Empower 3 - WKB25153

Article number: 25153

SYMPTOMS

  • The plate in position 1 of the sample manager is recognized but the plate in position 2 is not
  • The sample set begins and runs until it gets to the first vial in plate position 2
  • At that point in the sequence, the run fails and the "plate missing" error message appears

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • Sample Organizer
  • Empower 3

CAUSE

The user is loading the plate into position 2 by opening the door of the sample manager and placing it in there. They are not loading it via the sample organizer. 

FIX or WORKAROUND

Remove the plate from position 2 and load it through the sample organizer.

ADDITIONAL INFORMATION

id25153, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, SUP, UPCPORG, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

Not able to find a solution? Click here to request help.