ACQUITY pump error: Pump HW overpressure - WKB2495
Article number: 2495
SYMPTOMS
- Pump faults when starting flow
- Fault repeats on initialization and reset
- Fault repeats with pump heads removed
ENVIRONMENT
- Binary Solvent Manager (BSM)
- Quaternary Solvent Manager (QSM)
CAUSE
Bad transducer.
FIX or WORKAROUND
- Replace transducer (700002594) Transducer Assy, Head Mounted, 15 Kpsi
ADDITIONAL INFORMATION