ACQUITY UPLC PDA Detector leaks from the flow cell - WKB246816

Article number: 246816

SYMPTOMS

  • Leaking at the flow cell
  • Wavelength verification error
  • Shutter homing error

ENVIRONMENT

  • ACQUITY UPLC PDA Detector 
  • ACQUITY UPLC TUV Detector

CAUSE

Cracked or broken flow cell

FIX or WORKAROUND

Replace the flow cell. Refer to KB Article WKB191949 for the part numbers for the various flow cells

What are the part numbers for the various flow cells for ACQUITY UPLC Tunable UV and PDA detectors? - WKB191949 - Waters

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.