2695 Seal Wash Line Dry - WKB24682

Article number: 24682

SYMPTOMS

  • Seal-wash prime is initiated
  • The seal-wash pump comes on and the normal clicking sound is heard, but no flow moves through the seal-wash tubing
  • The seal-wash line is submerged in the seal-wash solvent bottle

ENVIRONMENT

  • Alliance 2695
  • Alliance e2695
  • Alliance 2795
  • Alliance HPLC System

CAUSE

The entire seal-wash line is dry. The seal wash pump isn't strong enough to pull the solvent through the line when totally dry.

ADDITIONAL INFORMATION