ACQUITY BSM shows "excessive leak back error" during dynamic leak test - WKB243155
Article number: 243155
SYMPTOMS
- Dynamic leak test fails to achieve set pressure for the accumulator
- Pressure increases and then falls back, primary pressure spikes, accumulator pressure rises, and cycle repeats
- Excessive leak back error eventually occurs
ENVIRONMENT
- ACQUITY UPLC BSM
- ACQUITY UPLC
CAUSE
Leaking accumulator transducer to vent valve compression fitting
FIX or WORKAROUND
- Replace the tubing (430001200 - Acc. A Transducer to Vent Valve, 430001199 - Acc. B Transducer to Vent Valve).
- Perform the dynamic leak test.
ADDITIONAL INFORMATION
A solvent droplet was visible on the tubing.