ACQUITY BSM shows "excessive leak back error" during dynamic leak test - WKB243155

Article number: 243155

SYMPTOMS

  • Dynamic leak test fails to achieve set pressure for the accumulator
  • Pressure increases and then falls back, primary pressure spikes, accumulator pressure rises, and cycle repeats
  • Excessive leak back error eventually occurs

ENVIRONMENT

  • ACQUITY UPLC BSM
  • ACQUITY UPLC

CAUSE

Leaking accumulator transducer to vent valve compression fitting

FIX or WORKAROUND

  1. Replace the tubing (430001200 - Acc. A Transducer to Vent Valve, 430001199 - Acc. B Transducer to Vent Valve). 
  2. Perform the dynamic leak test.

ADDITIONAL INFORMATION

A solvent droplet was visible on the tubing.

id243155, DLT, UPBINARY, UPBSM+, UPIBSM, UPIBSM+, UPPBSM