ACQUITY BSM randonly loses communication - WKB243375

Article number: 243375

SYMPTOMS

  • BSM loses communications both when idle and during an analytical run 
  • Module is not blank in console, but data frozen
  • Communication restored by hard reboot of the BSM, but happens again after

ENVIRONMENT

  • ACQUITY BSM
  • ALL ACQUITY MODULES

CAUSE

Faulty Ethernet cable

FIX or WORKAROUND

Replace the Ethernet cable and see if the issue is corrected over several analytical runs. 

ADDITIONAL INFORMATION