Skip to main content
Waters

With FTN, low system pressure when in inject position - WKB241372

Article number: 241372

SYMPTOMS

  • Pressure is normal in load position but much lower or even near zero in inject position

ENVIRONMENT

  • FTN sample manager

CAUSE

Leak at needle seal due to customer not recharacterizing after fitting new needle and seal

Issue could also stem from a damaged needle seal. Replace the needle seal and holder if required.

FIX or WORKAROUND

  1. Calibrate needle z axis.
  2. Characterize needle seal. (Pressure should now be normal in load and inject positions, and needle seal readiness test should now be successful.)

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.