Pressure fluctuations on an ACQUITY UPLC BSM - WKB24137
Article number: 24137
SYMPTOMS
- Pressure fluctuations on pump B
- ACQUITY BSM pressure fluctuates despite no detected leaks
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY UPLC BSM
- ACQUITY UPLC I-Class
- ACQUITY UPLC I-Class BSM
CAUSE
Faulty mixer
FIX or WORKAROUND
Replace the mixer
Check both aqueous and organic sides to ensure consistent pressure traces
ADDITIONAL INFORMATION
Due to higher compressibility, the issue is likely to be more prevalent on the aqueous side.
Mixers available:
700005341 ASSY, MIXER, 50uL 18K
205000854 ASSY MIXER, 100uL 18K
700002911 ASSY MIXER, 50uL 15K
205000404 ASSY, MIXER, 100UL, ZIRCONIA 15K psi
205000403 ASSY, FILTER MIXER, 425UL, 10K PSI
205000705 ACQUITY UPLC BSM, 380UL 15K PSI TFA MIXER