Peak response consistently higher or lower during analysis on ACQUITY UPLC SM-FTN - WKB241356
Article number: 241356
SYMPTOMS
- Reduced or saturated peaks when injecting same volume as on identical system
- Consistent areas are lower or higher than expected
ENVIRONMENT
- ACQUITY Arc Sample Manager FTN-R
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC M-Class
- nanoACQUITY UPLC
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
Sample syringe volume is incorrect.
FIX or WORKAROUND
Amend the sample syringe volume in the console to match the configuration of hardware.