ACQUITY needle wash drain tubing is detached from Y carriage - WKB241227
Article number: 241227
SYMPTOMS
- Tubing from Y carriage to BOS pump repeatedly detaches from the Y carriage
ENVIRONMENT
- ACQUITY Classic Sample Manager
CAUSE
The tubing end became baggy over time and no longer stays in place when pushed onto the barbed fitting.
FIX or WORKAROUND
- Trim a few mm of enlarged tubing from the end of the tube.
- Refit the tubing to the barbed fitting on the Y carriage.
- There is plenty of slack in the tubing to do this; new tubing is not required.
ADDITIONAL INFORMATION
If on inspection you discover the tubing has shrunk by about 25% in external diameter and length, the tubing (430001002) must be replaced.