No needle wash flows to waste when priming - WKB241261

Article number: 241261

SYMPTOMS

  • No needle wash flows to waste when priming

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Tubing from the solvent bottle into the solenoid valve is damaged.

FIX or WORKAROUND

  1. Ensure that there are no kinks in the tubing.
  2. Trim the ends of the tubing.
  3. Using a syringe, pull through solvent and place back into the solenoid valve.
  4. Confirm that solvent is flowing when priming.

ADDITIONAL INFORMATION

 