What are the wash solvents in PATROL? - WKB241200

Article number: 241200

ENVIRONMENT

  • PATROL UPLC System

ANSWER

Sample Wash = SNW (Strong needle wash)

  • Pushes sample through the needle, injection tower, and dilution loop and loads the sample into the injection loop; wash in contact with sample
  • Cleans the needle path, the dilution loop, and the injection valve

Diluent Wash Solvent = WNW (Weak needle wash)

  • Dilutes the sample and cleans the dilution loop
  • Requires miscibility with sample solvent; dilute the sample
  • Minimal wash cycles needed

External Needle Wash Solvent Considerations:

  • Clears excess sample from the injection tower
  • Especially with the online injection mechanism, requires miscibility, solubility with the sample