Skip to main content
Waters

Increased response, peaks in blanks - WKB241078

Article number: 241078

SYMPTOMS

  • Increased response for one or more analytes
  • Analyte is showing up in blanks
  • SM and BSM diagnostics pass

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC H-Class Bio
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class PLUS
  • ACQUITY UPLC H-Class PLUS Bio
  • ACQUITY UPLC I-Class PLUS
     
  • ACQUITY TQD system
  • TQ Detector
  • Xevo TQ MS
  • Xevo TQD
  • Xevo TQ-S
  • Xevo TQ-S micro
  • Xevo TQ-XS

CAUSE

Poor sample preparation with potentially contaminated solvents

FIX or WORKAROUND

  1. Prepare new samples or standards using all new solvent bottles. 
  2. If a new plate does not fix the issue, perform a system flush (Poor Inter-Instrument Reproducibility, Unstable Areas, High Baseline, Poor Peak Shape, or Unexpected Peaks - wkb234708), putting on new mobile phases afterwards.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.