Increased response, peaks in blanks - WKB241078
Article number: 241078
SYMPTOMS
- Increased response for one or more analytes
- Analyte is showing up in blanks
- SM and BSM diagnostics pass
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY TQD system
- TQ Detector
- Xevo TQ MS
- Xevo TQD
- Xevo TQ-S
- Xevo TQ-S micro
- Xevo TQ-XS
CAUSE
Poor sample preparation with potentially contaminated solvents
FIX or WORKAROUND
- Prepare new samples or standards using all new solvent bottles.
- If a new plate does not fix the issue, perform a system flush (Poor Inter-Instrument Reproducibility, Unstable Areas, High Baseline, Poor Peak Shape, or Unexpected Peaks - wkb234708), putting on new mobile phases afterwards.
ADDITIONAL INFORMATION
id241078, eluent, TQD, TQDIVD, UPBINARY, XEVOTQD, XEVOTQDIVD