Poor inter-instrument reproducibility, unstable areas, high baseline, poor peak shape, or unexpected peaks - WKB234708
Article number: 234708
SYMPTOMS
- Unstable peak areas for analyte or IS
- Running the plate on an identical system shows typical results
- High baseline
- Poor peak shape
- Potential analyte in blanks
- SM and BSM pass leak tests
- Potentially higher than expected system pressure
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
- ACQUITY UPLC I-Class
- ACQUITY UPLC H-Class PLUS
- ACQUITY UPLC H-Class PLUS Bio
- ACQUITY UPLC I-Class PLUS
- ACQUITY UPLC BSM
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
Contamination or blockage in the LC
FIX or WORKAROUND
- Perform a system flush using Magic Mix (equal parts ACN/MeOH/IPA/H2O with 0.1% formic acid).
- After the system flush, put on new mobile phases and columns.
ADDITIONAL INFORMATION
System Flush Protocol:
*Note: Be sure to place a union or a junk column on during the flush to make sure the column is not contaminated. Also, if possible, avoid touching the solvent sinkers when changing out the solvents.
- Place all mobile phase solvent lines, seal wash, weak needle wash, strong needle wash, and seal wash lines into a clean bottle containing 50% water/50% acetonitrile. NOTE: If there is an incompatibility between the mobile phase or solvents with the isopropanol, flush first with the appropriate intermediate solutions to ensure compatibility.
- Prime all mobile phase solvent lines for 5 min each.
- Prime the seal wash.
- Prime the Sample Manager (SM).
For a Fixed-Loop(FL) SM: Prime wash and sample syringes for 5 cycles.
For a Flow-Through Needle (FTN), prime wash for 60 seconds and prime sample syringe for 20 cycles. - Place mobile phase solvent lines, weak needle wash, strong needle wash, and seal wash lines into a clean bottle containing Magic Mix (25% water/25% methanol/25% isopropyl alcohol/25% acetonitrile/ 0.1% formic acid).
- Repeat steps 2-4.
- Set the fluidics valve on the mass spectrometer to the waste position and the column manager to bypass mode.
- Perform 15 injections from a vial or well containing Magic Mix.
SM: Run Time = 0.5 min / Flow Rate = 0.2 mL/min.
CM: Bypass
MS Method: Flow to Waste (Method Events), 0.5-Min Run Time
For a Fixed-Loop system: perform Full-Loop injections with 5x overfill.
For an FTN, perform 10-uL injections (or the maximum for your needle size). - Set the flow of the system at 0.2 mL/min using 50% A1 and 50% B1, or whatever two solvent lines you’d like to use, and allow to pump overnight.
- Repeat step 8 using 50/% A2 and 50% B2.
- Allow A2/B2 to flow for a minimum of 4 hours.
- Place lines in normal mobile phases and washes (using new bottles if possible) and replace the column.
- Repeat steps 2-4.
id234708, eluent, UPBINARY