Alliance leak during wet prime - WKB234650
Article number: 234650
SYMPTOMS
- Significant leak with no chromatography or pressure issues
ENVIRONMENT
- Alliance
CAUSE
Disconnected waste tubing from V3
FIX or WORKAROUND
Access the right-hand side of instrument and connect the tubing.
ADDITIONAL INFORMATION
When waste tubing becomes obstructed (squeezing waste lines excessively), back pressure can be sufficient to pull out waste tubing inside the instrument. As a consequence, all solvent from wet prime can flow onto the sample compartment.