Alliance leak during wet prime - WKB234650

Article number: 234650

SYMPTOMS

  • Significant leak with no chromatography or pressure issues

ENVIRONMENT

  • Alliance

CAUSE

Disconnected waste tubing from V3

FIX or WORKAROUND

Access the right-hand side of instrument and connect the tubing.

ADDITIONAL INFORMATION

When waste tubing becomes obstructed (squeezing waste lines excessively), back pressure can be sufficient to pull out waste tubing inside the instrument. As a consequence, all solvent from wet prime can flow onto the sample compartment.  