Cannot connect four eCords to CM-A and CM-Aux in UNIFI - WKB234213

Article number: 234213

SYMPTOMS

  • Error "Two eCords connected to column x" when more than two eCords are plugged into CM-A

ENVIRONMENT

  • UNIFI
  • Four columns
  • CM-A and CM-Aux

CAUSE

All four eCords are plugged into the CM-A positions.

FIX or WORKAROUND

Plug two eCords into the CM-A and two eCords into the CM-Aux.