"Sample Manager FTN: Column temperature range error" with UNIFI waters_connect - WKB230014
Article number: 230014
SYMPTOMS
- FTN with CM-A (no Column Heater)
- "Sample Manager FTN: Column temperature range error" appears after running system startup
- Error also produced during a run (3/4 into the run)
ENVIRONMENT
- waters_connect
- UNIFI Network
- ACQUITY Premier, ACQUITY I-Class
CAUSE
"Column Temperature" (for a CHA) was selected in the startup settings under "Sample Manager FTN", but system was configured to use CM-A. (See attached image.)
FIX or WORKAROUND
Clear the "Column Temperature" selection from system startup/Startup/Shutdown methods.