Set pressure diagnostics fails on ACQUITY UPLC M-Class - WKB229997
Article number: 229997
SYMPTOMS
- Set pressure diagnostics fails when testing up to pre-column
- Late retention times can be observed
ENVIRONMENT
- ACQUITY UPLC M-Class
CAUSE
A bad fitting for tubing connection from the flow controller to the Tee before column
or from vent valve to flow controller
FIX or WORKAROUND
Replace the appropriate piece of tubing. For tubing:
ADDITIONAL INFORMATION
If above tubing replacement doesn't fix the issue the vent valve is likely faulty and should be replaced