Alliance HPLC System gives "Column temp h/w fault (7)" error message - WKB229412
Article number: 229412
SYMPTOMS
- When setting the column temperature on the Alliance or setting up an inlet method in Empower/MassLynx, the error message "Column temp h/w fault (7)" appears
- Column heater/cooler power cable is plugged in and the on/off switch (on the lower-right side of the column oven) is in theOn position
- A different power cable was tested, but the error persists
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Faulty electric socket
FIX or WORKAROUND
- Plug the power cord into another electric socket.
- Test the unit at a desired temperature to verify that the error no longer occurs.
ADDITIONAL INFORMATION
id229412, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB