2414 data cannot be imported - WKB224264

Article number: 224264

SYMPTOMS

  • Cell temperature can be set from Empower
  • Can be set up and monitored
  • Chromatograms are captured during prep injection
  • When the sample is injected, the count starts but the chromatogram does not appear, and it times out after about 10 minutes

ENVIRONMENT

  • 2414 with IEEE connection
  • Empower

CAUSE

Poor contact of start signal cable

FIX or WORKAROUND

Disconnect and reconnect the start signal cable.

ADDITIONAL INFORMATION

id224264, 2414, 2414B, A-10RI