Rising back pressure and high delta on ACQUITY System - WKB224229

Article number: 224229

SYMPTOMS

  • Rising back pressure; pressure drops in cycles and increases again
  • High delta on ACQUITY BSM

    NOWTSZ 20211208 rising backpressure high delta.PNG

ENVIRONMENT

  • ACQUITY UPLC
  • ACQUITY UPC2 System

CAUSE

Partially clogged column

FIX or WORKAROUND

Replace the problematic column in the column compartment.

ADDITIONAL INFORMATION