How To turn Prep 150 LC System from preparative configuration to analytical - WKB221448

Article number: 221448

OBJECTIVE or GOAL

  • Change configuration of 2707 Sample Manager from Preperative to Analytical

ENVIRONMENT

  • 2707 Sample Manager
  • 2525/2545 Binary Gradient Module
  • 2535/2545/2555 Quaternary Gradient Module
  • Prep 150 LC System

PROCEDURE

  1. Install Analytical PM Kit (2010000196)
  2. Install Sample loop (standard 20uL 700000680 see WKB78006 for sizes
  3. Install correct sample tray, see WKB19842
  4. Install analytical flow cell WAS081140

ADDITIONAL INFORMATION

Installation of analytical components changes configuration from Prep 150 LC System to analytical HPLC

 

Not able to find a solution? Click here to request help.