Communication lost, blinking red light with Agilent 1260 Infinity DAD - WKB221359
Article number: 221359
SYMPTOMS
- Lost communication
- Blinking red light
- HPLC system modules unavailable ("grayed out") in Empower
- Module shows offline in Run Samples control panel
- Instrument failure
ENVIRONMENT
- Agilent 1260 Infinity
- Agilent LC
- G4212B 1260 DAD
- Agilent ELS
- Empower 3
CAUSE
- LAN cable connected incorrectly
- Computer unresponsive
FIX or WORKAROUND
- Power-down all Agilent HPLC components.
- Reboot the computer.
- Connect directly to LAC/E to "Start Windows normally".
- Power-up all Agilent LC components.
- Check system status. If communication is not restored, try steps 6 through 9.
- Remove the established chromatographic system in Configuration Manager.
- Click the "Tools" tab and configure new system through "Agilent PreConfiguration".
- Re-create the chromatographic system in Configuration Manager.
- Verify connection to the re-configured chromatographic system.
ADDITIONAL INFORMATION