Skip to main content
Waters

Communication lost, blinking red light with Agilent 1260 Infinity DAD - WKB221359

Article number: 221359

SYMPTOMS

  • Lost communication 
  • Blinking red light 
  • HPLC system modules unavailable ("grayed out") in Empower
  • Module shows offline in Run Samples control panel
  • Instrument failure

ENVIRONMENT

  • Agilent 1260 Infinity
  • Agilent LC
  • G4212B 1260 DAD
  • Agilent ELS
  • Empower 3

CAUSE

  • LAN cable connected incorrectly
  • Computer unresponsive

FIX or WORKAROUND

  1. Power-down all Agilent HPLC components.
  2. Reboot the computer.
  3. Connect directly to LAC/E to "Start Windows normally".
  4. Power-up all Agilent LC components.
  5. Check system status. If communication is not restored, try steps 6 through 9.
  6. Remove the established chromatographic system in Configuration Manager.
  7. Click the "Tools" tab and configure new system through "Agilent PreConfiguration".
  8. Re-create the chromatographic system in Configuration Manager.
  9. Verify connection to the re-configured chromatographic system.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.