Skip to main content
Waters

Baseline drifts upward with 2414 Refractive Index Detector - WKB221054

Article number: 221054

SYMPTOMS

  • Baseline drifts upward

ENVIRONMENT

  • 515 HPLC Pump
  • 2414 Refractive Index Detector
  • e2695 Alliance

CAUSE

  • Insufficient equilibration between the reference cell and analytical cell
  • Temperature
  • Contaminated system and column
  • Waste tubing plumbing
  • Mobile phase

 

FIX or WORKAROUND

  1. Insufficient equilibration between the reference cell and the analytical cell
    → Determine if improvement is possible by using a long purge time (about 30 minutes).

  2. Temperature (temperatures of the column heater and the 2414 detector cell)
    → Make the cell temperature as close as possible to the column temperature.

  3. Contaminated system and column
    → For column cleaning, refer to the Care & Use Manual for the column.
    See 2695 Separations Module System Cleaning Method-WKB91424 for system cleaning procedures.
    * Remove the solvent filter (the filter attached to the end of the line inserted into the solvent bottle) before cleaning the system, sonicate it, and reattach it after system cleaning.
    * During system cleaning, put the RI detector in the purge state to clean the reference cell as well.
    * During system cleaning, switch the valve of the manual injector to clean the valve as well.
    * During system cleaning, you can also monitor the baseline to check the drift condition.
    In that case, it is easy to drift without a column, so if you have a column that will no longer be used, you may want to connect it.

  4. Waste tubing plumbing
    → Determine if the waste line from the detector outlet is bent or if it is submerged in the waste in the solvent bottle.

  5. Mobile phase
    → Make fresh mobile phase and replace the entire bottle. When using a mix of multiple solvents, insufficient mixing may be the cause.

ADDITIONAL INFORMATION

id221054, 2414, 2414B, 515, A-10RI, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, eluent

Not able to find a solution? Click here to request help.