Baseline drifts upward with 2414 Refractive Index Detector - WKB221054
SYMPTOMS
- Baseline drifts upward
ENVIRONMENT
- 515 HPLC Pump
- 2414 Refractive Index Detector
- e2695 Alliance
CAUSE
- Insufficient equilibration between the reference cell and analytical cell
- Temperature
- Contaminated system and column
- Waste tubing plumbing
- Mobile phase
FIX or WORKAROUND
-
Insufficient equilibration between the reference cell and the analytical cell
→ Determine if improvement is possible by using a long purge time (about 30 minutes). -
Temperature (temperatures of the column heater and the 2414 detector cell)
→ Make the cell temperature as close as possible to the column temperature. -
Contaminated system and column
→ For column cleaning, refer to the Care & Use Manual for the column.
See 2695 Separations Module System Cleaning Method-WKB91424 for system cleaning procedures.
* Remove the solvent filter (the filter attached to the end of the line inserted into the solvent bottle) before cleaning the system, sonicate it, and reattach it after system cleaning.
* During system cleaning, put the RI detector in the purge state to clean the reference cell as well.
* During system cleaning, switch the valve of the manual injector to clean the valve as well.
* During system cleaning, you can also monitor the baseline to check the drift condition.
In that case, it is easy to drift without a column, so if you have a column that will no longer be used, you may want to connect it. -
Waste tubing plumbing
→ Determine if the waste line from the detector outlet is bent or if it is submerged in the waste in the solvent bottle. -
Mobile phase
→ Make fresh mobile phase and replace the entire bottle. When using a mix of multiple solvents, insufficient mixing may be the cause.
ADDITIONAL INFORMATION
id221054, 2414, 2414B, 515, A-10RI, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, eluent