Skip to main content
Waters

Chromatogram shows no peaks and low baseline - WKB221134

Article number: 221134

SYMPTOMS

  • No peaks seen
  • Low baseline

ENVIRONMENT

  • ACQUITY UV detector
  • Light guided flow cell

CAUSE

Flow cell is not allowing transmission of light.

FIX or WORKAROUND

  1. Clean the flow cell with 30% orthophosphoric acid, 0.5 mL/min for 60 minutes.
  2. If sample energy does not increase, replace the flow cell.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.