Can a Column Manager - Active be used to direct flow and divert to waste? - WKB21662

Article number: 21662

ENVIRONMENT

  • ACQUITY ELS Detector
  • ACQUITY Column Manager - Active (CM-A)

ANSWER

Yes. Configuring the Column Manager - Active in advanced mode of operation will allow command of eight-position, nine-port switching valves independently. This is a custom configuration that requires the following:

  1. Configure the operating mode for the "Column Manager - Active" in the software console.
  2. Program the column manager method to command the valve positions.
  3. Plumb the fluidics according to need.
    • Choose a switching valve that will operate as the inlet from the sample manager. This valve will be responsible for commanding the column position.
    • Choose a switching valve that will operate at the outlet from the column. This valve will act as a divert valve for commanding flow between waste and detector.

ADDITIONAL INFORMATION

 

 

id21662, A-30MCM, UPCMA, UPCMAIVD, UPELS, UPELSARC, UPPCMA

 