How much sample do I put in the vials for an Alliance 2695? - WKB2135
Article number: 2135
ENVIRONMENT
- Alliance 2695
ANSWER
It depends on the vial used. Refer to the Waters Sample Vials and Accessories Brochure.
Also see:
What does "residual volume" mean when referring to autosampler vials?
ADDITIONAL INFORMATION
For example, if using the standard 2-mL flat-bottom vial, the residual volume of this vial for an Alliance 2695 is 750 uL. This means that if you do five injections of 10 uL from the same vial, you need (5 x 10(uL)) + 750(uL) = 800 uL in the vial.