Skip to main content
Waters

Error: x axis home h/w fault - Sample Manager - WKB2139

Article number: 2139

SYMPTOMS

  • Console shows error X axis H/W fault
  • Grinding noise when sample manager tries to home after a reset
  • Restart clears the error, but faults again during injection

ENVIRONMENT

  • ACQUITY UPLC Sample Manager

CAUSE

Unit has not been characterized in a long time.

FIX or WORKAROUND

  1. Remove the two plates.
  2. Power-off the Sample Manager.
  3. Move the needle carriage gently back and forth couple times, paying attention to the flex cable to avoid damaging it.
  4. Power-on the Sample Manager.
  5. Put the plates back in place.
  6. Prime wash solvents.
  7. Characterize needle seal.
  8. Characterize needle and sample loop volumes.

ADDITIONAL INFORMATION


id2139, audible noise, UPBSMIVD, UPIBSMIVD, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPSMIVD, UPSMPMGR, UPSMPMGR

Not able to find a solution? Click here to request help.