Article number: 209636

SYMPTOMS

  • SM-FTN Seal force sensor h/w fault (10.0) error for sample manager FTN during initialization, just after the needle carriage touches the rotary tray
  • Error is intermittent
  • After load cell replacement, z-axis calibration out of range error observed, followed by inability to calibrate needle seal

ENVIRONMENT

  • ACQUITY UPLC SM-FTN

CAUSE

Faulty sensor interface PCB

FIX or WORKAROUND

Replace the sensor interface PCB (700005195).

ADDITIONAL INFORMATION