Skip to main content
Waters

With MassLynx, PDA does not stop at the run end time - WKB209672

Article number: 209672

SYMPTOMS

  • All other ACQUITY modules stop at run end and enter idle state, but the PDA carries on running until it is reset
  • Chromatogram stops at run end time; data trace is still collecting in the console

ENVIRONMENT

  • MassLynx
  • ACQUITY

CAUSE

Unknown

FIX or WORKAROUND

  1. Reload ICS and firmware for the detector.
  2. Cycle power to the detector and the PC when the load is complete.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.