Skip to main content
Waters

2487 detector keeps rebooting - WKB2095

Article number: 2095

SYMPTOMS

  • Unit starts the boot up sequence, lights the lamp and goes through the 5 minute warm up then reboots (keeps initializing) 

ENVIRONMENT

  • 2487 Dual-Wavelength Absorbance Detector

CAUSE

Battery voltage is too low

FIX or WORKAROUND

Replace the battery Part Number: WAT080443

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.