Frequent ACQUITY UPLC "communication failure" error messages - WKB209495

Article number: 209495

SYMPTOMS

  • Frequent "communication failure" errors in ACQUITY Console for all instrument modules
  • New PC configuration

ENVIRONMENT

  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class
  • ACQUITY UPLC M-Class
  • ACQUITY UPLC
  • Empower
  • MassLynx
  • SCIEX Analyst

CAUSE

The Windows Firewall is killing ACQUITY communications.

FIX or WORKAROUND

  1. Disable Windows Firewall for the instrument LAN card (see WKB18188).
  2. Verify that instrument communications are stable. 

ADDITIONAL INFORMATION

 